5.2.1
(c) Loosen three screws of the D
2
lamp cover and
separate this cover (remove lead wires one by one
through the cover).
Cover
Fastening screw
(d) Loosen the 3 screws fastening the D
2
lamp by turning
each screw 2 revolutions with a blade-edged
screwdriver.
Fastening
screw
(e) Pull out the D
2
lamp while turning it clockwise.
(f) Fit a new D
2
lamp and turn it counterclockwise until
there is no gap between the mounting hole and the
fastening screw.
Then, tighten the 3 fastening screws evenly in an
alternate order with the blade-edged screwdriver.
<Good example>
<Bad example>
No gap allowable
D
2
lamp
Mounting hole
Fastening
screw
5 - 5
Содержание CHROMASTER 5160