APPENDIX 2. PLACE FOR INSTALLATION
2.1 Installation Conditions
(1) Power
Supply
Line voltage: 24 V DC (supplied from the PC board at the
rear of pump unit)
(2) Installation
Location
In Chromaster 5110 pump.
2.2 Ambient Conditions
(1) Operating Temperature 4 to 35
°
C
For measurement under the most stable condition, it is
recommended to install this product in an air-conditioned
room at 20 to 25
°
C.
(2) Operating Humidity 25 to 85%
(3) Atmospheric
Gas
(a) Adequate ventilation is required.
(b) The room shall not be filled with highly corrosive gas of
acid or alkali.
(4) Other General Precautions
(a) Shall not be exposed to direct sunlight.
(Performance degradation or coating discoloration
might be caused.)
(b) Shall be free from strong vibration or shock which can
be felt by your body.
(Fine adjustment of mechanism might be lost.)
(c) Shall be free from a heating device such as gas burner,
electric heater or oven in the vicinity that may heat the
main unit cover (beyond 70
°
).
APPENDIX 2 - 1
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...