3 - 51
PROG No.=1
PRESSURE MAIN
MAX
MIN
(MPa)
20.0 0
(a)
TIME
FLOW
EVENT
0.0
1.000
10.0
1.000
11 21 31 41
10.1
1.000
20.0
1.000
(b) 25.0
10 20 30 40
30.0
2.000
30.1
1.000
40.0
1.000
00
92
Fig. 3-4 Example of Program
(a) For EVENT, up to 4 event codes are settable at each
time point. It is possible to generate contact outputs
at each EVENT terminal at the same time. You can
also obtain an internal action, such as a pump stop
and the ringing of a buzzer.
Example)
EVENT= 1 - 1 - ENT - 2 - 1 - ENT
(b) In case that an event setting is desired independent of
FLOW, only the event code showed be entered.
Press ENT to skip all the entries you do not want until
you reach event column.
In the SET PROGRAM mode, the interpolation of mixing ratio,
flow rate, etc. that occurs in a time program is as follows.
Parameter Flow
Rate
Interpolation Linear
T
N
T
N+1
T
N
, T
N+1
: Preset time
Fig. 3-5 Interpolation of Mixing ratio and Flow rate
NOTE
:
For generating a stepwise gradient program, utilize
T
N+1
- T
N
= 0.1.
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...