3.2.3 Setting of Error Out Contact
Specify whether or not to output an error contact signal to
another instrument upon occurrence of an error in the column
oven. When the other instrument receives the error contract
signal, its operation is stopped.
UTILITY (LEAK SENSOR=1, ANALOG OUT=2,
3 ERROR OUT=3, CALIBRATION=4)
ERROR OUT SIGNAL (DISABLE=0, ENABLE=1)
1
3 ENT
Set whether or not to output the
error contact signal.
1 ENT
Monitor screen
SET UP
UTLTY
NOTE
: Error contact output will always be issued regardless of
the setting of ERROR OUT contacts for only leak error
occurrence.
3.2.4 Setting of Temperature Calibration
If there is a deviation between the indicated temperature of the
Chromaster 5310 column oven and the actual temperature, the
deviation can be corrected by setting for temperature calibration.
For instance, if the actual temperature measured by a calibrated
thermometer is 40.5
°
C while the indicated temperature of the
column oven is 40.0
°
C, the offset value for the deviation is +0.5,
so make the setting as illustrated below.
3 - 13
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...