3 - 29
3.2.9 Link Program Function
Plural programs can be linked.
NOTE:
When connecting with CDS, setting should be set to
NO. In this case, setting will be made on the system
manager side.
(1) Setting of Link Program
Set the conditions for link program. Explained below is an
example where you set one run of program No. 1 at the
first step, conditioning time at 30 minutes, 3 runs of
program No. 2 at the second step and conditioning time at
20 minutes.
Monitor screen
UTLTY
(SET UP)
UTILITY
(
CONDITION=1
,
UNIT=2
,
FLUSH=3
5 EVENT=4
,
LINK=5
,
OTHERS=6
)
LINK PROGRAM
(
YES=1, NO=0
)
1
LINK RESTART
(
YES=1, NO=0
)
0
1 ENT
Monitor screen
After executing a link program by a few steps, execution of the next steps will be
avoided and the link program will return to the first step and will be executed
again. (Assume that a 5-step link program is set. If you set link restart after
the 2nd step in this case, the 3rd step will not be executed and execution will
restart from the 1st step.)
Also, when plural cycles of the same step are set, the link program will return to
the first step and will be executed so far as the link restart is set, though all the
set cycles are not completed. Setting should be made when a pump is set in
the initial status ([IN] status is indicated at the top right of LCD). Setting is
impossible during gradient elution and conditioning.
0 ENT
If manual setting screen is displayed, press [INIT] to obtain the
programming mode.
Select LINK (LINK PROGRAMA)
5 ENT
Select YES
1 ENT
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...