(ii) Check the position and useful life of the xenon
lamp.
(b) When peak separation is not good
(i)
Check the distilled water.
* Distilled water should be used soon after its
purification. If left for a long time in a container
made of resin or the like, fluorescent substance
may mix in with the distilled water.
(ii) Check for bubbles
If water is directly injected into an empty flow cell,
small bubbles may remain in the cell. To
prevent this, first inject ethanol or the like and
then water.
(iii) Change the method of spectrum measurement
(via time program).
Measure the Raman spectrum of water using a
time program. But first check the fluctuation of
FL value via wavelength shift.
Set the excitation(Ex) wavelength at 350 nm and
emission(Em) wavelength at 450 nm using
the
WL
key.
Then zero the FL value with the
A/Z
key.
Next, with the
WL
key, check the Em
wavelength near 397 nm where the FL value is
maximum, and the Em wavelength near 380 nm
where the FL value is minimum.
The minimum FL value should be less than half
the maximum FL value.
4 - 12
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...