NOTE
: When an input value of VOLUME
exceeds the syringe capacity set as
SYRINGE VOL in UTILIITY, an error of
[CHECK TOTAL VOL.] will occur.
NOTE
: The syringe moves to the specified
volume position. The syringe valve
position will be switched over
automatically according to the syringe
movement.
To the port3: aspiration movement.
To the port4: discharge movement.
NOTE
: When it is discharge movement, wash
fluid in the syringe will be discharged to
the port4 side of the syringe valve.
In this case, remove the closed cap in
port4 of the syringe valve before the
SYR command is carried out. Without
removing, it reduces the lifetime of the
syringe.