3.3 GLP-Related Functions
3.3.2 Logbook Setting (Valve)
When using the optional selector valve, information on the
replacement date of the valve seal (a maintenance item) and
number of valve switchings after the replacement is recorded in
logbook form.
Enter the year when valve
seal was replaced.
2004 ENT
Monitor screen
Carry out setting for the
logbook in which
maintenance information
is recorded.
2 ENT
VALVE1 SEAL CHANGE (MM DD YYYY)
<2000 - 2100> 9 1 2010
VALVE1 SEAL CHANGE (MM DD YYYY)
<1 - 12> _9 1 2010
Enter the month when v
seal was replaced.
6 ENT
alve
VALVE1 SWITCHING LAST CHANGED DATE
23 times
9/1/2010
Monitor screen
ENT or ESC key
Check the logbook information.
1 ENT
Enter the day when valve
seal was replaced.
2 ENT
VALVE1 SEAL CHANGE (MM DD YYYY)
<1 - 31> 9 _1 2010
LOGBOOK (REPORT=1, RESET=2)
2
1 ENT
GLP
GLP (LOGBOOK=1, KEY LOCK=2)
1
3 - 16
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...