NOTE
: If the monitor display screen does not appear on the
LCD within 3 minutes after turning on the power, or if
an error message appears on the LCD, refer to “5.6
Troubleshooting” or “5.7 Error Messages”.
1)
Turn on the power switch of the detector.
2)
Liquid Crystal Display (LCD) presents the initial screen,
which changes to the initializing screen after a few seconds.
At the same time, a test for self-diagnosis (initialization) is
carried out.
* The current channel setting LED and the Xe LAMP LED
will light in blue.
<Initial screen>
FL Detector 8928148-00
© Hitachi High-Technologies Corp.2010
↓
<Initializing screen>
WL DRIVE & LAMP INITIALIZING
The sequence of wavelength drive (check of reference point
for drive), lighting of the lamp and auto zeroing is performed.
3)
The first monitor screen will be indicated after initialize-
processing is completed normally.
Press
△
▽
key to indicate the second and the third
monitor screen.
Press
△
▽
ESC key to return to the first monitor
screen.
* An error message will be presented if an abnormality
occurs during initializing.
Check Section 4.6.
3 - 2
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...