*3) Higher sample solubility solvent is used.
(Some types of samples are dissolved with 100% organic
solvent.)
NOTE
: Stainless steel (SUS316), Vespel®, fluorine resin,
polypropylene (PP), ethylene-propylene rubber (EPDM)
and perfluoroelastomer materials are used for the wetted
parts of the instrument. Never use solvents that would
corrode these materials.
2.3.3 Sample Injection Procedure
Cut Injection Method Procedure
The cut injection method, the injection volume, the lead
volume, the rear volume, and the air volume are restricted by
the following relational expression. A stroke of the syringe
capacity can’t be used all.
<175 µL of Standard syringe >
vi + vl + vr + va
≦
165 µL
vi :Injection volume
(The sample net volume to the flow path)
vl :Lead volume
(The disposal volume of the lead part of a sample)
vr :Rear volume
(The disposal volume of the rear part of a sample)
va :Air volume
(The air volume of the lead and rear part of a sample)
100 µL loop is used for the autosampler as the standard.
In general, the lead volume and the rear volume are set as 30
μ
l each when using 10
μ
l of sample.
2 - 13
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...