WARNING
Electric shock due to improper grounding
This instrument is designed to meet the basic insulation
stipulated in the international standard IE C61010-1.
To prevent an electric shock hazard due to improper
grounding, provide a proper grounding connection.
(a)
Be sure to use the grounded 3P power cable, which
is supplied as a standard accessory for the
instrument. The use of a different type of power
cable may result in an electric shock hazard.
Connect the 3P power cable to a grounded 3P power
outlet as a rule.
(b)
If a grounded 3P power outlet is not available, use a
3P table tap with a grounding wire or a 3P-2P
adapter. In this case, be sure to provide proper
grounding connection.
For grounding connection, use a screw having a
diameter of M4 or larger and a turn count of 3 or
more in threading, and a wire having a thickness of
1.25 mm.
(1) Grounding
Terminal
Shall have a grounding construction for a ground resistance
of 100
Ω
or less.
APPENDIX 1 - 4
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...