3 - 53
1
Program number can be from 1 to 9.
2
When you want to define a new program, enter 1.
As a result, you will delete the previous program.
For modifying the contents of program, enter 0.
3
If pump pressure rises beyond the level set here, the pump
automatically stops.
Set pressure can be from 0.1 to 60.0 MPa.
4
If pump pressure falls below the level set here, the pump
automatically stops.
Set pressure can be from 0.0 to 60.0 MPa.
5
The set time can be from 0.0 to 600.0.
6
Flow rate valid entries are from 0.00 to 5.000
7
Input the event code. At the set time point, you can stop
the pump, generate an audible signal, markers, timers, etc.
When properly interfaced, these signals can be output to an
external system. No display will show on the screen
unless an event is set. When an event is not set, press only
the ENT key.
8
Repeat
steps
5
to
7
until you complete the program.
9
Program generation will be terminated by pressing ESC
and the monitor screen will reappear.
For confirming the input conditions, check the parameters at each
time point by using
keys.
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...