5 - 18
(4) Fill the ultrasonic cleaner bath with water. Put the beakers in
the ultrasonic cleaner bath and perform ultrasonic washing
for 5 to 6 minutes.
(5) Spread a large piece of filter paper on the table. In the
following order, take out the valve chest, ball, and valve seat
from the beaker. Use tweezers to avoid contamination.
Then assemble the check valve using the correct
procedure.
NOTE
: When assembling, do not make a mistake in the check
valve orientation and placement of the valve seat faces.
The glossy side of the valve seat should face the ball.
NOTE
: The valve seat and ball are used as a pair. Be sure to
assemble the same pair together.
Valve holder
Packing (A)
Check valve
Packing (A)
Check valve
Packing (B)
Packing (A)
Check valve
Packing (A)
Check valve
Packing (B)
Valve
holder
Suction
Valve
Beaker
Beaker
Beaker
Discharge
Valve
Beaker
Beaker
Beaker
Check valve holder
Beaker
Ultrasonic cleaner
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...