APPENDIX 2 DESCRIPTION ON CONTACT SIGNAL COMMUNICATION
1. Outline
The following two types of contact signals are used in the
Chromaster series.
(1) Contact signals (START, ERROR, BUSY) entering the
e-Line connector.
These are contact signals having input/output functions that
are used in common for each unit. These signals are
respectively formed from the START (IN/OUT), ERROR
(IN/OUT), and BUSY (IN/OUT) lines.
(2) Individual contact signals used by each unit.
These are contact signals that are input and output via 3-pin
connector. Their names and functions vary with the unit.
A name that indicates the function is given at each
connector.
•
Example of output contact signal: EVENT of pump
•
Example of input contact signal: LAMP OFF IN of detector
2. Contact Signals Entering the e-Line Connector
Each unit has two e-Line connectors for connection at the rear of
the main unit, and the signals of each unit in series via the e-Line
cable, synchronism of the Chromaster series system can be
easily achieved.
For obtaining synchronism with the L-2000 series and/or D-2500
by way of contact signals, use the cables listed below for the
connection.
APPENDIX 2 - 1
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...