solvent before use. For measurement, select a
wavelength having less absorption.
It is recommended to use a solvent specifically
prepared for high-speed liquid chromatography.
(b) Be sure to be deaerated when an organic solvent is
used.
Ultrasonic
cleaner
Aspirator
Absorbance
Wavelength
(4) Cleaning of Micro-syringe
(a) After and before sampling, wipe off the needle of a
micro-syringe.
(b) Before aspiring a wash fluid, wipe off the needle and
wash the inside of the syringe repeatedly, i.e., seven
times or more.
(c) After injection, wash the syringe immediately.
(5) Fastening of Connection Tube
If the Teflon tube is overtightened, its inside diameter is
made smaller to cause an abnormal increase in pressure,
3 - 76
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...