4.1 How to Check the Specifications and Performance
Temperature sensor
35mm
Detailed view
Fig. 4-4 Attachment of Temperature Sensor
NOTE
: In cases where a non-specified temperature
sensor is used as a substitute for the specified
one, be sure to measure a surface temperature in
the vicinity of the groove of the preheat base.
If any other temperature is measured, accurate
temperature data may not be attained.
(3) Attach the inside cover. And before closing the front door of
the column oven, run the wires of the temperature sensor
through the piping connection opening.
(4) Set the terminal of the thermocouple to the thermometer,
and set the column oven temperature to 40
°
C.
(5) When the column oven temperature indication has
stabilized, measure the thermocouple temperature.
Specification: Difference between temperature measured by
thermocouple and the column oven temperature
indication should be within
±
1.0
°
C.
4 - 6
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...