3 - 31
(2) Actions in Link Program
In a link program, two or more programs are used in
succession.
Explanation is given in case a link program is set under the
following conditions.
(a) Programs
PROGRAM No. = 1
PRESSURE VALUE : MAX
MIN
MPa
19.6
1.0
TIME
CONCENTRATION
FLOW
EVENT
(
min
)
%A %B %C %D
(
mL
/
min
)
1 2 3 4
0.0
10
90
0
0 1.000
5.0
10
90
0
0
10.0
20
80
0
0
15.0
20
80
0
0
20.0
10
90
0
0
PROGRAM No. = 2
PRESSURE VALUE : MAX
MIN
MPa
19.6
1.0
TIME
CONCENTRATION
FLOW
EVENT
(
min
)
%A %B %C %D
(
mL
/
min
)
1 2 3 4
0.0
30
70
0
0 1.000
5.0
40
60
0
0
10.0
40
60
0
0
15.0
30
70
0
0
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...