Set a photomultiplier voltage.
Press numeric key (1 to 5, usually input 3) and then the
ENT key.
* Default value: 3
(4) Photomultiplier voltage setting is now completed and the 1
st
monitor screen returns.
3.2.5 Other
settings
Set the following items.
•
Em
BANDWIDTH
•
ERROR
OUT
•
LEAK
SENSOR
•
LAMP
(1) Em BANDWIDTH setting
A bandwidth value of the emission monochromator is
settable. Standard is 15 nm and wide setting is 30 nm.
(a) Press
the
UTLTY
SET UP
key.
(b) The following display will be presented.
<Utility initial screen>
UTILITY (OFFSET=1, TIME CONSTANT=2,
_5
USE PROG=3, PMT VOLT=4, OTHERS=5)
Select OTHERS.
Press numeric key 5 and then the ENT key.
(c) The following display will be presented.
<Others setting screen>
OTHERS (Em BANDWIDTH=1, ERROR OUT=2,
1
LEAK SENSOR=3, LAMP=4)
Select Em BANDWIDTH.
Press numeric key 1 and then the ENT key.
3 - 12
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...