Select EXCHANGE STANDARD.
Press numeric key 4 and then ENT key.
(4) The following display will be presented.
<Standard Xe lamp lifetime setting screen>
XE LAMP EXCHANGE STANDARD (1 - 2500) H
_500
The standard lifetime of the Xe lamp is settable here.
Press numeric key (to input 1 to 2 5 0 0) and then the ENT
key.
* Check the part number of your Xe lamp and input the
correct value listed below.
Part No.
Part Name
Guaranteed Life
Limit
Allowable Input
650-1500
150 W Xe lamp
150 hours
500 hours
500 or less
890-2019
Long-life Xe lamp
500 hours
1,000 hours
1000 or less
* Default value: 500 hours
(J851152 Xe lamp is attached with shipment.)
(5) The setting of a standard Xe lamp lifetime is now completed
and the following display will be presented.
<Setup initial screen>
SETUP: DETECT THERMO CONTRAST S. NO.
CH1
USE
3 02AB-111
* When you press the ENT key here, the setup items
selecting screen is presented to allow reentry.
* The selected setting will be fetched upon turning on
power supply at the next time.
6)
Turn OFF the power switch.
3 - 51
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...