3.2.1
Set an offset value.
Press [numerical] key (within [0] to [1][0][0][0]: Usually input
[0]), and then the [ENT] key.
* Settable range : 0 to 1000 AU, settable in increments
of 1
* Default
value : 0
■
Since the negative voltage input range is around -10 mV
with the usual data processor, a baseline lower than
-10 mV level may disable integration of some
chromatogram peaks (Fig. 3-2(a)). In this case, the
offset function adjusts the baseline, which is shifted to
the negative side, to a signal level within the input range
of the data processor (Fig. 3-2(b)).
(b) Offset Used
(a) Offset Not Used
-10
(corresponding to
-10 mV)
Fig. 3-2 Example of Offset Usage
4)
Upon completion of offset setting, the first monitor screen
reappears.
Although the input numeral is not indicated at ABS, “+*” will
be suffixed to ABS value when a numeral other than 0 is
input.
First Monitor Screen
0.0 250
0.0000+
*
1
D2&W
TIME
WL
ABS
PROG
LAMP
[I]
3 - 11
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...