(3) Put a new injection port seal into the injection port.
Attach it to the injection valve, and tighten it up with your
hand.
Turn the taper part of the injection
port seal to the injection side when
attach it.
Injection port seal
Taper part
Injection port
Fig. 5-3
Attachment of Injection Port Seal
NOTES
: 1. Turn the taper part of the injection port seal to
the injection port side when attach it.
2.
Tighten the injection port with your hand until
it is finger-tight.
3.
Avoid using tools to prevent the injection port
seal from deformation.
(4) After replacing the injection port seal, press the
WASH
key
to carry out injection port cleaning.
NOTES
: 1. After replacement of the injection port seal U,
make sure liquid is not leaking through the
seal.
2. If the tip of needle is deformed, the service
life of port seal may be significantly
shortened. Use the needle free from tip
deformation.
5 - 5
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...