(ii) Rack parameter setting procedure
The rack parameter is settable in a numerical
input method or in the robotic mode where the
needle is actually moved for adjustment.
2 ENT
1 ENT
Input the code of the rack which parameter is to be set.
Select “INPUT VALUE” for numerical input or “ROBOTIC” for
the robotic mode.
1 ENT
Input the X1 coordinate.
57.1 ENT
Input the X2 coordinate.
178.6 ENT
Input an Nx value.
10 ENT
Input the Y1 coordinate.
0.0 ENT
RACK CODE (1-16)
1
OPERATION MODE (INPUT VALUE=1
1 ROBOTIC=2)
X1 (0.0-180.0) mm
57.1
X2 (0.0-180.0) mm
178.6
Nx (1-40)
10
Y1 (0.0-155.0) mm
UTILITY (SEQ=1, RACK=2, WASH PARAM=3,
1 OTERS=4
0.0
(to be continued)
3 - 26
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...