3.3
3.3 Error Output
(1) Error
Output
If the vacuum pump of 6-channel degassing unit is in either
status below, an error will be output and the vacuum pump
will stop. At the time of error output, the monitor lamp of
this degassing unit lights in orange.
(a) When the internal pressure of degassing chamber has
not reached 11 kPa at 10 minutes after start of the
vacuum pump drive.
(b) When vacuum level has exceeded 13 kPa.
(c) At occurrence of pressure sensor error
NOTE
: When an error output is issued in the 6-channel
degassing unit, analysis cannot be started from
CDS.
(2) Release of Error
Error status can be released by turning on power supply to
the Chromaster 5110 pump again.
NOTE
: If the error recurs despite restart of the pump, be
sure to consult your dealer or nearest Hitachi
High-Technologies service representative.
3 - 3
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...