3.6.2
3.6.2 Characteristics of Organic Solvents
Characteristics of organic solvents as mobile phase for liquid
chromatograph are shown in Table 3-3. Use it for reference.
WARNING
Ignition of Flammable Chemicals
•
This instrument is not explosion-proof. In unattended
operation, do not use organic solvents having an ignition
point below 70
°
C.
•
Beware of ignition hazard when using flammable
chemicals such as organic solvents.
(a) Do not bring a heat or flame source near the
instrument.
(b) Well-ventilate the laboratory room where the
instrument is used.
(c) Always check the following conditions. If an
abnormality is found, stop operation immediately.
Leakage of solvent or waste solution.
Leakage of solvent inside the instrument.
•
When using flammable chemicals, be careful about
possible ignition due to static electricity.
To prevent the build-up of static electricity, use a
conductive container for waste solution and provide
proper grounding connection to it.
WARNING
Explosion of Vapor from Flammable Chemicals
If a flammable chemical such as organic solvent leaks from
the flow path of the instrument and its vapor concentration
exceeds the explosion limit, it may cause spontaneous
combustion with dangerously explosive results.
When using a flammable and volatile chemical, be sure to
check for leakage from the instrument flow path and
ventilate the laboratory room adequately.
3 - 25
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...