(2) Sequential program setting
(a) Setting
method
As a method of setting a sequential program, the
setting procedure is shown below on the following
conditions; analysis time 5 minutes and injection of
120 samples of 10
μ
L each sequentially once.
STEP RACK VIAL VOL(µL) INJ TIME [I]
1 1 1 10.0 1/1 0.0 [C]
This display appears at the end of initialization after power-
on.
PROG
EDIT MODE (MODIFY=0, NEW=1)
0
Select MODIFY for editing the previous program or NEW for
creating a new program.
0 ENT or 1 ENT
STEP RACK VIALS VOL(µL) INJ TIME
1 1 1- 1 10.0 1 5.0
Input a rack code.
1 ENT
STEP RACK VIALS VOL(µL) INJ TIME
1 1 1- 1 10.0 1 5.0
Input the first vial number.
1 ENT
STEP RACK VIALS VOL(µL) INJ TIME
1 1 1- 1 10.0 1 5.0
Input the last vial number.
120 ENT
STEP RACK VIALS VOL(µL) INJ TIME
1 1 1-200 10.0 1 5.0
Input an injection volume (
μ
L).
10 ENT
(to be continued)
3 - 13
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...