3.2 Utility Setting
(d) Setting the ON/OFF Status of D
2
Lamp/W Lamp
1) Press key.
SET UP
UTLTY
2) The utility setup screen appears.
Utility Setup Screen
Select OTHERS.
Press [numerical] key [5], and then the [ENT] key.
3) The OTHERS setup screen appears.
Select LAMP.
Press [numerical] key [3], and then the [ENT] key.
4) The LAMP setup screen appears.
LAMP ON/OFF (OFF=0, D2=1, W=2, D2&W=3)
1
OTHERS (PROCESSOR=1, ERROR OUT=2
1
LEAK SENSOR=3, LAMP=4, GAIN=5)
UTILITY (OFFSET=1, TIME CONSTANT=2,
1
USE PROG=3, POLARITY=4, OTHERS=5)
Set the ON or OFF status of each lamp.
Press [numerical] key (within [0] to [3]), and then the
[ENT] key.
* Default value: 3
For selection of D2 lamp : Press [numerical] key [1],
and then the [ENT] key.
For selection of W lamp : Press [numerical] key [2],
and then the [ENT] key.
For selection of D2 and
W lamps
: Press [numerical] key [3],
and then the [ENT] key.
For selection of lamp
OFF status
: Press [numerical] key [0],
and then the [ENT] key.
3 - 18
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...