Operating suggestions are indicated below.
Cut Injection Method
When applying the cut injection method, the loop
capacity more than the injection volume is required.
Select loop capacity within the area indicated in the
following figure.
The sample introduction volume is selected within the
shadow area of the following figure. (For example, 200
µL loop is required when measuring 100 µL sample.
This means that the sample volume requires the double
loop capacity.)
0
10
20
30
40
50
60
70
80
90
100
20
40
60
80
100 120 140 160 180 200
Loop Capacity [µl]
S
am
ple
Inj
ect
io
n
Vo
lum
e
[
µl]
Fig. 2-7 Cut Injection Method
All Volume Injection Method
When applying all volume injection method, the loop
capacity much more than the injection volume is required.
Select loop capacity within the area indicated in the
following figure.
The sample introduction volume is selected within the
shadow area of the following figure. (For example, 200
µL loop is required when measuring 40 µL sample. This
means that the sample volume requires quintuple loop
capacity.)
2 - 10
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...