4.2 Confirmation of Performance Specifications
4.2 Confirmation of Performance Specifications
4.2.1 Items of Performance Specification Check
(1) Measurement of Light Source Lamp Energy (See 4.4.1)
•
Measuring wavelength: 250 nm
•
Reference at the time of installation
Energy indication of 150,000 or more at wavelength
250 nm
The indicated energy value has a slight difference among
the light source lamps due to their characteristics.
(2) Measurement of Wavelength Accuracy (See 4.4.1)
Check wavelength accuracy using the bright lines of D
2
lamp at 486 and 656 nm, and the bright line of Hg lamp at
254 nm.
Specification:
Wavelength accuracy of each bright line
shall be within
±
1 nm.
(3) Measurement of Baseline
•
Conditions........................ Room temperature constant,
air in flow cell, time constant
2.0 s,
Noise in data sampling period
400 ms
Drift in data sampling period
1600 ms
•
Measuring wavelength..... 250 nm
•
Specifications .................. Noise
: 0.5
×
10
-5
AU or less
Drift : 1.0
×
10
-4
AU/h or
less
(when measured at
least 1 hour after
power-on)
4 - 2
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...