Table 3-1
Reagents for Mobile Phase that may be Restricted
Use is
Possible
Use is Possible if
50% or Lower
Use is Possible if 10% or
Lower
Reagents Unsuitable for
SUS316
Phosphoric
acid
Sodium
phosphate
Acetic acid
Ammonium citrate
Ammonium nitrate
Citric acid
Butyric acid
Sodium nitrate
Disodium phosphate
Ammonium formate
Ammonium perchlorate
Ammonium phosphate
Boric acid
Formic acid
(up to pH3)
Hydrochloric acid
(up to pH3)
Potassium nitrate
Sodium hydrogencarbonate
Sodium carbonate
Ammonium chloride
Potassium chloride
Sodium chloride
Tolufluoroacetic acid
Note 1:
Max. concentration of
above reagents should
be 5-6%.
Note 2:
When the above
reagents have been
used necessarily, wash
all the flow paths with
distilled water completely
without fail.
Do not use reagents that corrode SUS316 though they are not
shown in the table above. If viscosity of solvent is high,
pumping is impossible.
3 - 55
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...