3 - 34
⑦
Delay time (initial)
Liquid is delivered in the initial status of the first step (initial
condition of program No. 1 in this example). When “start”
is not input during the delay time, column flushing in
⑧
will begin.
When “start” is input during the delay time, the system
returns to action in
②
.
⑧
Column flushing
(ii) Without start input after gradient elution
The encircled numerals in this execution example
correspond to the step numbers of screen display.
TIME %A %B %C %D
FLOW PRESS
[
IN
]
0.0 10 90 0 0 1.000 5.5 C
1
TIME %A %B %C %D
FLOW PRESS
[
FL
]
0.0 10 90 0 0 1.000 5.5 C
1
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...