background image

 

 

 
 
 
 
 
 

Softcontrol Data/F 

 

Manual 

 
 
 
 
 
 

 

 
 
 
 
 

version 2.0 

 
 
 

Содержание Softcontrol Data/F

Страница 1: ...Softcontrol Data F Manual version 2 0...

Страница 2: ......

Страница 3: ...ing 19 10 2 Switch on the device 19 10 3 Control unit and display in measuring mode 20 10 4 Parameterization 25 10 4 1 Password entry 25 10 4 2 Parameters and settings 26 10 5 LAN interface 29 10 5 1...

Страница 4: ...4...

Страница 5: ...rst softener column must be monitored Monitoring of the second softener column when connected in series is not permitted Characteristics Monitoring of the soft water hardness Option to parameterize th...

Страница 6: ...uction processes of the user their malfunctions failures or damage or consequential da mage that are directly or indirectly connected to the monitoring or control circuit of the device or that have or...

Страница 7: ...ordance with the in stallation regulations Operation is only permitted in rooms with a relative humi dity of 20 to 60 and ambient temperatures of 5 C to 50 C It must be en sured that no condensation c...

Страница 8: ...of the sample inputs Before removing the sensor from the measuring chamber the power supply to the device must be interrupted Otherwise water may splash out of the measuring cham ber There is constant...

Страница 9: ...off valves regulate the extraction of minimal process water quantities and the subsequent pressure free drainage of the water sample Additional reagents and consumables are not required The measuring...

Страница 10: ...ed on an USB stick Softcontrol Data F can be connected to a LAN via an Ethernet interface The parameterization and control of the device function is possible via the Ethernet interface Softcontrol Dat...

Страница 11: ...n Provided that the measurement pause has been set to 0 minutes in the parameter menu inter val the soft water measurements are carried out as long as a flow is detected The flow is recognized and eva...

Страница 12: ...m next to the device as well as a pressure free drain with nominal diameter 5 8 DN 15 in the immediate vicinity below the device For the extraction of the raw and soft water there must be branches wit...

Страница 13: ...supplied screw in connection into the shut off branch of the soft water and tightly plug the sample hose OD 4 mm into the screw in connection on both sides and into the straight plug connector black...

Страница 14: ...y unit The electrical connections relevant for the user are located on the underside of the controller card The incoming and outgoing signal lines are connected to the terminal blocks on the underside...

Страница 15: ...ntial free fault message limit value alarm user 20 Switching output 1 potential free fault message limit value alarm user 21 Switching output 2 potential free fault message limit value sensor alarm us...

Страница 16: ...16 9 3 2 Overview of electrical connections USB Display connection service alarm limit value alarm water meter SALIS external alarm offline contact release measurement 4 20mA current loop LAN...

Страница 17: ...er failure or error fault detected See also chapter Overview external alarms open External signaling of limit value alarms via potential free contact Output limit value alarm Terminals 19 and 20 The f...

Страница 18: ...free normally open contact Terminals 31 and 32 External requirements do not connect any external voltage potential free external contact necessary Installation water meter See also chapter Operating...

Страница 19: ...green Online LED A short cyclical flashing of the green LED shows that the internal control functions are active and Softcontrol Data F is working properly The specified process parameters must be se...

Страница 20: ...ated display In addition there are three LEDs for operating and fault signaling in the upper area Button functions 4 Buttons Display LED display green LED readiness display flashes cyclicyally goes ou...

Страница 21: ...ll in an unstable measuring state Soft water concentration greater than raw water concentration caused by impermissible brine entry 2 line calibration status calib Raw and soft water switch valves dev...

Страница 22: ...easured value below the set limit value of 2 ppm SW 2 ppm measured value above the set limit value of 2 ppm SW 2 ppm possibly brine in the soft water similarly also output of limit value alarms Note T...

Страница 23: ...hapter Service alarm active S internal function test carried out result determined sensor quality too low Note For causes see chapter Service alarm active calib The device carries out a function test...

Страница 24: ...ta on USB stick Note USB stick must be plugged into an external PC If possible reformat this stick or exchange it for an identical USB stick from OFS GmbH offline Softcontrol Data F is switched in off...

Страница 25: ...confirmed with the Menu OK button 10 4 1 Password entry The password is 07580 The buttons have the following functions for entering the password Confirm password Change the numerical value of the acti...

Страница 26: ...e chapter operating mode alarm delay Delay of the alarm in the range of 1 30 minutes in steps of 1 minute 10 minutes measure Selection of the unit of measurement between dH mmol l f or ppm dH language...

Страница 27: ...USB stick network access ON activation or OFF deactivation of the network Factory setting OFF Writing the network parameters on an USB stick To write the current network parameters to the USB stick th...

Страница 28: ...ters from a USB stick the USB stick with the NETWORK TXT file must first be inserted Reading is started with the Menu OK button The parameters read in IP address gateway subnet mask and port for commu...

Страница 29: ...t the USB stick to the controller card 2 Call up the menu item Write CONFIG TXT A file with the name CONFIG TXT is created on the USB stick This contains the relevant network parameters Example factor...

Страница 30: ...command 5 PARAM3 Parameter according to command 6 PARAM4 Parameter according to command 7 Reserve Currently not used The following command codes are defined code command description parameter 0 GET_DA...

Страница 31: ...ters The status word contains the coded representation of the current operating status of the device If the corresponding bit is set the corresponding entry is active The status word is coded as follo...

Страница 32: ...0 3 Raw water hardness 11 232 Raw water hardness 12 0 Soft water hardness 13 SW 2 34 ppm 13 0 Soft water hardness 14 0 Soft water hardness 15 13 Soft water hardness 16 0 Soft water limit value 20 3 60...

Страница 33: ...It only makes sense to start this functional test if the hardness value of the soft water is below the specified limit value the soft water is free of brine and there is sufficient water pressure in...

Страница 34: ...on the flow measuring chamber at the same time pull the sensor straight out 5 Pull the protective cap straight off the new sensor taking care not to touch or damage the electrodes Note In the event of...

Страница 35: ...number given on the packaging and on the delivery note is recommended The sensor should be stored in the original packaging at 10 C to 15 C and used as soon as possible The protective cap must remain...

Страница 36: ...he factory default settings the following steps must be carried out 1 Pull out the mains plug 2 Press the button and simultaneously 3 Plug in the power plug and keep both buttons pressed until the str...

Страница 37: ...brine entry into the soft water The soft water is to be checked for brine using a comparative conductivity measu rement with the raw water Basically the rinsing times of the softener should be extend...

Страница 38: ...losed valve Remove the hoses from the measuring chamber and check whether drops of water are escaping despite the closed valves replace affected valve Low brine concentration in the soft water Rinse t...

Страница 39: ...surement result raw and soft water tension M Data of measurement Measurement result soft water tension O Offline status ON OFF Q Functional test Raw soft and mixed water tension sensor quality Example...

Страница 40: ...7 CL1794 SC961 SS1533 PF257916 2 Copy the new software version sent by the manufacturer into a directory on the Windows PC 3 Switch off Softcontrol Data F or disconnect from the power supply 4 Remove...

Страница 41: ...ound in the same directory in which the DfuSe software was installed After successful driver installation the system must have added a new device called STM Device in DFU Mode A successful driver inst...

Страница 42: ...ed 12 When the software is shown in the field of view the update can be started by clicking on the Upgrade field The following message is confirmed by clicking on Yes and the up date process starts 13...

Страница 43: ...is ready for operation with the new software 15 Device accessories 4 m flexible pressure proof PA line 4 mm OD 2 mm ID sampling hose 2 m flexible pressure proof PA line 15 mm OD 11 5 mm ID drainage ho...

Страница 44: ...o the valid German drinking water ordinance germ count 5000 CFU ml Water hardness range raw water 70 ppm 1070 ppm or adequately in mmol l f or ppm Water hardness range soft water 0 2 ppm 18 ppm or ade...

Страница 45: ...gen For verification in accordance with Low Voltage Directive 2014 35 EU the following standards were applied EN 61010 1 2010 Zum Nachweis entsprechend EMV Richtlinie 2014 30 EG wurden folgende Normen...

Страница 46: ...OFS Online Fluid Sensoric GmbH Heidelbergweg 9 07580 Ronneburg Germany manual Softcontrol Data F version 2 0 issue 01 2021 technical changes reserved...

Отзывы: