background image

 
 

 

 

34 

 

 

11.3.

 

Test sensor contact 

 
After opening the device door and removing the sensor cable from the sensor, the alarm must 
be  signaled  within  20  seconds  by  the  red  LED  lighting  up,  "no  sensor"  being  shown  in  the 
display and the external fault signal "changing sensor" being switched over. When the sensor 
cable is plugged into the sensor, the alarm must be cleared within 10 seconds. 

 
11.4.

 

Test external signaling  

 
For this test, the service relay is de-energized so that the relay contact opens. This fault case 
can be simulated simply by pulling the mains plug, which also enables the external signaling of 
a power failure on the device. The switching states of the relay  are explained in  the section 
"Electrical installation". 

 

11.5.

 

Sensor - installation / exchange / storage / conditions of use 

 
Sensor installation / sensor exchange 
 

The replacement of the sensor is to be carried out as follows: 

1.

 

Pull out the mains plug and open the appliance door. 

2.

 

Pull the sensor out of the holder and turn it about 45 ° towards you. 

3.

 

Loosen the screw connection on the sensor connector and pull the sensor connector off 
the sensor. 

4.

 

Press down the pressure ring on the flow measuring chamber, at the same time pull the 
sensor straight out. 

5.

 

Pull the protective cap straight off the new sensor, taking care not to touch or damage 
the electrodes. 

Note: 

In the event of a warranty check, close the old sensor tightly with the protective cap 

and send it back to the folding box. The liquid should not run out of the protective cap.

 

6.

 

Insert the new sensor straight into the plug-in system of the flow chamber and press it 
in firmly as far as it will go. The sensor must click into place audibly and tangibly. 

7.

 

Check the sensor for a tight fit by pulling it gently. 

8.

 

Insert the plug of the sensor cable into the sensor socket. 

Attention

: observe reverse polarity protection 

The plug connection must be made without any great effort. Otherwise it 
can be assumed that the electrical connection is faulty due to bent contacts 
in the plug.  

9.

 

Fix the sensor cable to the sensor with a screw connection and press the installed sensor 
system into the holder on the mounting plate 

10.

 

Close  the  appliance  door  and  plug  the  safety  plug  back  into  the  mains.  Softcontrol 
Data/F checks the sensor plug connection when it is restarted and reports "no sensor" if 
the sensor is not recognized. In this case, Softcontrol Data/F must be switched off again 
and the plug connection to the sensor checked for contact security. 

Содержание Softcontrol Data/F

Страница 1: ...Softcontrol Data F Manual version 2 0...

Страница 2: ......

Страница 3: ...ing 19 10 2 Switch on the device 19 10 3 Control unit and display in measuring mode 20 10 4 Parameterization 25 10 4 1 Password entry 25 10 4 2 Parameters and settings 26 10 5 LAN interface 29 10 5 1...

Страница 4: ...4...

Страница 5: ...rst softener column must be monitored Monitoring of the second softener column when connected in series is not permitted Characteristics Monitoring of the soft water hardness Option to parameterize th...

Страница 6: ...uction processes of the user their malfunctions failures or damage or consequential da mage that are directly or indirectly connected to the monitoring or control circuit of the device or that have or...

Страница 7: ...ordance with the in stallation regulations Operation is only permitted in rooms with a relative humi dity of 20 to 60 and ambient temperatures of 5 C to 50 C It must be en sured that no condensation c...

Страница 8: ...of the sample inputs Before removing the sensor from the measuring chamber the power supply to the device must be interrupted Otherwise water may splash out of the measuring cham ber There is constant...

Страница 9: ...off valves regulate the extraction of minimal process water quantities and the subsequent pressure free drainage of the water sample Additional reagents and consumables are not required The measuring...

Страница 10: ...ed on an USB stick Softcontrol Data F can be connected to a LAN via an Ethernet interface The parameterization and control of the device function is possible via the Ethernet interface Softcontrol Dat...

Страница 11: ...n Provided that the measurement pause has been set to 0 minutes in the parameter menu inter val the soft water measurements are carried out as long as a flow is detected The flow is recognized and eva...

Страница 12: ...m next to the device as well as a pressure free drain with nominal diameter 5 8 DN 15 in the immediate vicinity below the device For the extraction of the raw and soft water there must be branches wit...

Страница 13: ...supplied screw in connection into the shut off branch of the soft water and tightly plug the sample hose OD 4 mm into the screw in connection on both sides and into the straight plug connector black...

Страница 14: ...y unit The electrical connections relevant for the user are located on the underside of the controller card The incoming and outgoing signal lines are connected to the terminal blocks on the underside...

Страница 15: ...ntial free fault message limit value alarm user 20 Switching output 1 potential free fault message limit value alarm user 21 Switching output 2 potential free fault message limit value sensor alarm us...

Страница 16: ...16 9 3 2 Overview of electrical connections USB Display connection service alarm limit value alarm water meter SALIS external alarm offline contact release measurement 4 20mA current loop LAN...

Страница 17: ...er failure or error fault detected See also chapter Overview external alarms open External signaling of limit value alarms via potential free contact Output limit value alarm Terminals 19 and 20 The f...

Страница 18: ...free normally open contact Terminals 31 and 32 External requirements do not connect any external voltage potential free external contact necessary Installation water meter See also chapter Operating...

Страница 19: ...green Online LED A short cyclical flashing of the green LED shows that the internal control functions are active and Softcontrol Data F is working properly The specified process parameters must be se...

Страница 20: ...ated display In addition there are three LEDs for operating and fault signaling in the upper area Button functions 4 Buttons Display LED display green LED readiness display flashes cyclicyally goes ou...

Страница 21: ...ll in an unstable measuring state Soft water concentration greater than raw water concentration caused by impermissible brine entry 2 line calibration status calib Raw and soft water switch valves dev...

Страница 22: ...easured value below the set limit value of 2 ppm SW 2 ppm measured value above the set limit value of 2 ppm SW 2 ppm possibly brine in the soft water similarly also output of limit value alarms Note T...

Страница 23: ...hapter Service alarm active S internal function test carried out result determined sensor quality too low Note For causes see chapter Service alarm active calib The device carries out a function test...

Страница 24: ...ta on USB stick Note USB stick must be plugged into an external PC If possible reformat this stick or exchange it for an identical USB stick from OFS GmbH offline Softcontrol Data F is switched in off...

Страница 25: ...confirmed with the Menu OK button 10 4 1 Password entry The password is 07580 The buttons have the following functions for entering the password Confirm password Change the numerical value of the acti...

Страница 26: ...e chapter operating mode alarm delay Delay of the alarm in the range of 1 30 minutes in steps of 1 minute 10 minutes measure Selection of the unit of measurement between dH mmol l f or ppm dH language...

Страница 27: ...USB stick network access ON activation or OFF deactivation of the network Factory setting OFF Writing the network parameters on an USB stick To write the current network parameters to the USB stick th...

Страница 28: ...ters from a USB stick the USB stick with the NETWORK TXT file must first be inserted Reading is started with the Menu OK button The parameters read in IP address gateway subnet mask and port for commu...

Страница 29: ...t the USB stick to the controller card 2 Call up the menu item Write CONFIG TXT A file with the name CONFIG TXT is created on the USB stick This contains the relevant network parameters Example factor...

Страница 30: ...command 5 PARAM3 Parameter according to command 6 PARAM4 Parameter according to command 7 Reserve Currently not used The following command codes are defined code command description parameter 0 GET_DA...

Страница 31: ...ters The status word contains the coded representation of the current operating status of the device If the corresponding bit is set the corresponding entry is active The status word is coded as follo...

Страница 32: ...0 3 Raw water hardness 11 232 Raw water hardness 12 0 Soft water hardness 13 SW 2 34 ppm 13 0 Soft water hardness 14 0 Soft water hardness 15 13 Soft water hardness 16 0 Soft water limit value 20 3 60...

Страница 33: ...It only makes sense to start this functional test if the hardness value of the soft water is below the specified limit value the soft water is free of brine and there is sufficient water pressure in...

Страница 34: ...on the flow measuring chamber at the same time pull the sensor straight out 5 Pull the protective cap straight off the new sensor taking care not to touch or damage the electrodes Note In the event of...

Страница 35: ...number given on the packaging and on the delivery note is recommended The sensor should be stored in the original packaging at 10 C to 15 C and used as soon as possible The protective cap must remain...

Страница 36: ...he factory default settings the following steps must be carried out 1 Pull out the mains plug 2 Press the button and simultaneously 3 Plug in the power plug and keep both buttons pressed until the str...

Страница 37: ...brine entry into the soft water The soft water is to be checked for brine using a comparative conductivity measu rement with the raw water Basically the rinsing times of the softener should be extend...

Страница 38: ...losed valve Remove the hoses from the measuring chamber and check whether drops of water are escaping despite the closed valves replace affected valve Low brine concentration in the soft water Rinse t...

Страница 39: ...surement result raw and soft water tension M Data of measurement Measurement result soft water tension O Offline status ON OFF Q Functional test Raw soft and mixed water tension sensor quality Example...

Страница 40: ...7 CL1794 SC961 SS1533 PF257916 2 Copy the new software version sent by the manufacturer into a directory on the Windows PC 3 Switch off Softcontrol Data F or disconnect from the power supply 4 Remove...

Страница 41: ...ound in the same directory in which the DfuSe software was installed After successful driver installation the system must have added a new device called STM Device in DFU Mode A successful driver inst...

Страница 42: ...ed 12 When the software is shown in the field of view the update can be started by clicking on the Upgrade field The following message is confirmed by clicking on Yes and the up date process starts 13...

Страница 43: ...is ready for operation with the new software 15 Device accessories 4 m flexible pressure proof PA line 4 mm OD 2 mm ID sampling hose 2 m flexible pressure proof PA line 15 mm OD 11 5 mm ID drainage ho...

Страница 44: ...o the valid German drinking water ordinance germ count 5000 CFU ml Water hardness range raw water 70 ppm 1070 ppm or adequately in mmol l f or ppm Water hardness range soft water 0 2 ppm 18 ppm or ade...

Страница 45: ...gen For verification in accordance with Low Voltage Directive 2014 35 EU the following standards were applied EN 61010 1 2010 Zum Nachweis entsprechend EMV Richtlinie 2014 30 EG wurden folgende Normen...

Страница 46: ...OFS Online Fluid Sensoric GmbH Heidelbergweg 9 07580 Ronneburg Germany manual Softcontrol Data F version 2 0 issue 01 2021 technical changes reserved...

Отзывы: