3.2 UTLTY Setting
3.2.2 Analog Temperature Output
Set the output temperature for the analog output connector on
the rear of the instrument. Output temperature conforms to the
relational expression given below.
Temperature [
°
C] = (output voltage [mV]/9.6) - 10
SET UP
UTLTY
2 ERROR OUT=3, CALIBRATION
=
4)
ANALOG OUTPUT TYPE (OVEN =1, AMBIENT=2)
Select an output temperature.
1 ENT
Monitor screen
1
2 ENT
UTILITY (LEAK SENSOR=1, ANALOG OUT=2,
3 - 12