Wavelength calibration is conducted by entering a
new offset value (value to which the deviation is
added).
Press numeric keys (from 1 to 2 0) and then the
ENT key.
*
Input range for offset value: 1 to 20
* Calculation of new offset value:
New offset value = present offset value –
wavelength deviation
Example: 12 = 10 – (-2)
(iii) The
1
st
monitor screen appears.
(b) Calibration of emission wavelength
(i) Press
the
UTLTY
key, and the utility setting
screen appears.
<Utility setting screen>
UTILITY (OFFSET=1, TIME CONSTANT=2,
98
USE PROG=3, PMT VOLT=4, OTHERS=5)
Press numeric keys 9
8 and then the ENT
key.
(ii) The emission wavelength calibration mode
screen appears.
<Emission wavelength calibration mode screen>
Em WL ADJUST (1 - 20)
10 (ONLY FOR FACTORY USE) (-2)
The present offset value is indicated at Em WL
ADJUST. Indicated at the bottom right of the
screen is the wavelength check result (deviation
of wavelength; indication range of –10 to 10).
Wavelength calibration is conducted by entering a
new offset value (value to which the deviation is
added).
Press numeric keys (from 1 to 2 0) and then the
ENT key.
* Input range for offset value: 1 to 20
* Calculation of new offset value:
New offset value = present offset value –
wavelength deviation
Example: 12 = 10 – (-2)
4 - 17
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...