4.3 Periodic Checkup
(2) Cleaning of Leak Sensor
<Item to be prepared>
Hexagonal wrench (P/N: N315003)
(a) Remove the leak sensor fastening screw with the
furnished hexagonal wrench and separate the leak
sensor from the drain tray.
(b) Wipe off the water adhering to the leak sensor and
drain tray. Then, fasten the leak sensor to the drain
tray again.
NOTE
: When washing the leak sensor, be careful not to
damage the leak sensor tip. If damaged, the
leak sensor may not work normally or an error
may occur.
Leak sensor
fastening screw
Fig. 4-2 Leak Sensor Fastening Screw
NOTE
: Mount the leak sensor at the correct position.
If the mounting position is inappropriate or the
leak sensor is damaged, the sensor may not work
normally or an error may occur.
4 - 8
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...