(12) [When the plunger wash is applied]
Wash the pump plunger with the wash fluid2.
Switch the 3-way solenoid valve to the wash fluid2 side.
Switch the syringe valve to the plunger wash side (port 4).
Deliver the wash fluid2 to the pump plunger with the wash
pump (specified time).
After washing, it moves to the idle status.
Buffer tube
Injection valve
Syringe valve
Syringe
Wash pump
5
1
3
2
4
Plunger
wash
Needle
Injection port
5
4
3
2
1
From Pump To Column
6
Waste
solution
Loop
Waste
solution
Wash port
Sample
Rack
Wash
fluid2
P
3-way solenoid valve
Degassing
unit
Wash
fluid1
2 - 25