3.2 UTLTY Setting
UTILITY (LEAK SENSOR=1, ANALOG OUT=2,
4 ERROR OUT=3, CALIBRATION=4)
OFFSET (
°
C)
ENT or ESC key
Monitor screen
0.5
4 ENT
Select either plus or minus
side for deviation.
1 ENT
SET +OFFSET (0.0 - 5.0)
°
C
0.5
Set an offset value for the
correction.
0.5 ENT
SELECT MODE (+OFFSET=1, -OFFSET=2)
1 OFFSET
(
°
C)=-1.0
Present setting is
indicated.
SET UP
UTLTY
NOTE
: 1. For conducting temperature calibration, be sure to
use a calibrated thermometer for measuring
temperature.
2. There is a slight difference between the set and
actual temperatures depending on the temperature
setting. In temperature calibration, it is
recommended to determine the deviation at the set
temperature that is used most frequently.
3 - 14
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...