5 - 20
5.9 Parts Replacement for the Plunger Wash Mechanism
<Required Articles>
Hexagon wrench
(P/N: N315005, Size: B4)
Workshop tools
(P/N: 810-1353)
Double-end wrench
(P/N: N311017, Size: 13/17)
Flathead screwdriver
(P/N: N330062)
Receiving tray
Wash mechanism seal (P/N: 885-1884x 2)
Plunger seal(Pump seal)(P/N; 891-1213x 2)
*Replace it as necessary.
Wash bottle
Ultrasonic cleaner
(1) Disconnect the inlet tube and outlet pipes (1)(2) from the
pump heads (Refer to 5.3 Replacement of Pump Seal).
(2) Loosen the hexagon screws alternately with a hex wrench
and remove them.
(3) Carefully pull out the pump head toward you with both
hands.
When the plunger seal remains in the pump head,
replace it if needed (Refer to 5.3 Replacement of Pump
Seal).
When the plunger seal remains in the plunger, replace it
when the wash mechanism plunger guide is removed.
Pump head
Spring washer
Hexagon screws
Hex wrench
Plunger wash mechanism
Guide
Wash mechanism plunger guide
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...