7.2 Connection of Column Data Tags
When the switching valve is mounted, the column management
system updates the history data of the columns used in the
analysis, interlocked with the switching valve, each time an
analysis is completed, suspended and brought into an abnormal
end.
To ensure that the history data of the columns used in the
analysis are properly updated, connect the column data tags to
each appropriate slot according to the figure below.
The tag connected to an inappropriate slot will prevent correct
history data management.
A-1
A-2
A-3
G-1
G-2
G-3
Appended Fig. 1-22 Slot numbers
(1) Where No Switching Valve is Used
Where the column management system is used in the
Chromaster 5310 column oven which is set not to use a
switching valve, the data on the tags which are connected
to the A-1 slot (for analysis column) and the G-1 slot (for
guard column) as shown in the figure above will be updated.
No data on the tags connected to other slots will be
updated.
APPENDIX 1 - 26
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...