4 - 5
4.1.3 Flow Rate Precision Check
Check the precision of flow rate in the following procedure.
(1) Set the following conditions to autosampler.
STEP RACK VIALS VOL (uL) INJ TIME
1 1 1- 1 10.0 6 3.0
Press the ESC key to complete setting.
(2) Prepare methanol for the mobile phase and detergent.
Wait until the pump, detector, etc. are warmed up and the
baseline stabilizes.
(3) Press the WASH key to clean the needle and injection port,
and fill the flow path of the syringe system with the cleaning
agent.
(4) Place a vial containing at least 1 mL of standard sample in
rack position No. 1 and set the rack in the autosampler.
(5) Press the START key of autosampler to start analysis.
The analysis is carried out 6 times. The operating
conditions during analysis are indicated on the display.
Ensure beforehand that the room temperature will not
change suddenly during analysis.
(6) When the analysis is completed, the initial screen appears
and stops. If ALL END is set, ALL END screen appears and
stops. If you press the IDLE key, the initial screen appears.
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...