3.2 Utility Setting
3.2.4 Setting the Polarity
When NEGATIVE is selected, absorbance can be output in a
negative value.
SET UP
1)
Press key.
UTLTY
2)
The utility setup screen appears.
Utility Setup Screen
1
USE PROG=3, POLARITY=4, OTHERS=5)
UTILITY (OFFSET=1, TIME CONSTANT=2,
Select POLARITY.
Press [numerical] key [4], and then the [ENT] key.
3)
The polarity setup screen appears.
Polarity Setup Screen
1
POLARITY (POSITIVE=1, NEGATIVE=2)
Press [numerical] key ([1] or [2]: Usually input [1]), and then
the [ENT] key.
* Default value: 1
4)
Upon completion of polarity setting, the first monitor screen
reappears.
When NEGATIVE is selected, absorbance value becomes
negative and “-ABS“ appears on the monitor screen.
0.0 250 0.0000
1 D2&W
TIME
WL
-ABS
PROG
LAMP [I]
3 - 14
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...