3.3 GLP-Related Functions
3.3.2 Key
Lock
Keys on the keyboard are settable in the unacceptable status.
Wrong key input is prevented by applying key lock to the
operation panel.
NOTE
: When the GUI controller or CDS is connected,
operation with the UI pad becomes ineffective.
1)
Press key.
2)
The GLP setup screen appears.
GLP Setup Screen
Select KEY LOCK.
Press [numerical] key [2], and then the [ENT] key.
3)
The KEY LOCK setup screen appears.
GLP
KEY LOCK (YES =1, NO =0)
0
GLP (LOGBOOK=1, KEY LOCK=2,
1
D2 CHECK=3, W CHECK=4, Hg CHECK=5)
For validating
keyboard operation : Press [numerical] key [0], and then
the [ENT] key.
For invalidating
keyboard operation : Press [numerical] key [1], and then
the [ENT] key.
Key lock is released in the following cases;
i) When the ESC (LOCK OFF) key has been pressed.
ii) At occurrence of an error.
3 - 26
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...