(2) Notes on Handling Samples that Contain a High
Concentration of the Compound of Interest
When the concentration of the compound of interest is
relatively high, various error factors will arise.
The greatest factor is that the excitation beam is
substantially absorbed near the entrance slit of a cell and it
cannot reach the cell center adequately.
An extreme situation is shown in Fig. 3-8. Fluorescence is
emitted near the entrance slit of a cell, but it will not be
taken into the emission monochromator.
Excitation beam
Fluorescence emitted
Fluorescence
Fig. 2-8 Sample Having Extremely High Concentration
2 - 12
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...