<Scan monitor screen (spectrum acquisition under way)>
SCANNING SPECTRUM NO.: 1 MODE: FL (Em)
Ex: 350 nm Em: *** nm
(v) When the recording of spectral data finishes, the
1
st
monitor screen returns automatically.
(e) Setting of recorder full scale range, speed and memory
No.:
(i)
Press the RECORD key and the recorder setting
screen appears.
<Recorder setting screen>
RECORD (RCD RANGE=1, RCD SPEED=2,
1
SPECT
RCD=3)
Press numeric key 1 and then the ENT key.
(ii) The recorder full scale setting screen appears.
<Recorder full scale setting screen>
RCD RANGE (1 - 1000)
1000
Press numeric keys 1 0 0 0 and then the ENT
key.
(iii) The
1
st
monitor screen returns.
(iv) Press the RECORD key. And the recorder
setting screen appears.
<Recorder setting screen>
RECORD (RCD RANGE=1, RCD SPEED=2,
2
SPECT
RCD=3)
Press numeric key 2 and then the ENT key.
(v) The recorder speed setting screen will appear.
<Recorder speed setting screen>
RCD SPEED (40 mm/min=1, 60 mm/min=2)
2
4 - 7
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...