(continued)
STEP RACK VIALS VOL(µL) INJ TIME
1 1 1-120 10.0 1 5.0
Input the number of injections from each vial.
1 ENT
STEP RACK VIALS VOL(µL) INJ TIME
1 1 1-120 10.0 1 5.0
Input a cycle time (min).
5 ENT
STEP RACK VIALS VOL(µL) INJ TIME
2
Repeat input by the same method as for STEP1.
When pressing the
ESC
key, the monitor screen reappears.
ESC
STEP RACK VIAL VOL(µL) INJ TIME [I]
1 1 1 10.0 1/1 0.0 [C]
(b) STEP deleting method
Unnecessary steps in a sequential program can be
deleted.
STEP RACK VIALS VOL(µL) INJ TIME
3 _1 1- 1 10.0 1 5.0
Display the unnecessary step and press the
DEL
key.
STEP RACK VIALS VOL(µL) INJ TIME
- 1 1- 1 10.0 1 5.0
Press the
ENT
key.
STEP RACK VIALS VOL(µL) INJ TIME
3 1 5- 5 10.0 1 5.0
The old contents of the relevant step are deleted and those
of the next step are brought in.
NOTE
: STEP No. 1 cannot be deleted.
3 - 14
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...