2.4 Specifications
(d) Number of steps storable
Up to 100 steps in a total of nine files
(15) GLP Functions
Recording of the changeover number of times and
exchange dates of the optional changeover valve.
(16) Other Function
Liquid leak sensor
Temperature control is stopped upon detection of liquid
leak.
(17) Options
(a) GUI
controller
(b) UI
pad
(c) Selector
valves
One of the following two kinds of optional valves is
selectable.
(i) 3-column selector valve
Model............................. RV500-100
(IDEX Health&Science
LLC, USA)
Wetted part.................... SUS316, PEEK
Pressure tightness......... 35 MPa
Solvents not allowed ..... Concentrated sulfuric
acid and concentrated
nitric acid are not
applicable.
(ii) 6-way 2-position valve
Model............................. RV700-100
(IDEX Health&Science
LLC, USA)
Wetted part.................... SUS316, PEEK
Pressure tightness......... 35 MPa
Solvents not allowed ..... Concentrated sulfuric
acid and concentrated
nitric acid are not
applicable.
(d) Column management system
2 - 12
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...