From the pump
Fig. 4-10 Removing Air Bubbles
NOTE
: Applying too much pressure may damage the flow cell.
Set the pressure limiter of the pump at 1.0 MPa max.
before removing air bubbles.
4.4.2 Disassembly and cleaning of flow cell
(1) Disassembly of Flow Cell
If contamination is not removed by just flushing the interior
of the flow cell, then disassemble and clean the cell in the
procedure below.
Shielding plate
Retaining screw 1
Retaining screw 2
Setscrew 1
Setscrew 2
Fig. 4-11 Flow Cell Unit
(a) Loosen retaining screws 1 and 2 and detach the
shielding plate.
(b) Detach the cell window.
4 - 26