5 - 9
5.4 Replacement of Purge Valve Seal
CAUTION
Injury of Hand by Seal remover PV
When removing a seal from the Manual Purge Valve, the
seal remover of PV can injure your hand or finger.
Carefully read this instruction manual before starting
operation, and use care with Seal remover PV.
<Required articles>
Seal remover PV (P/N: 893-1285)
Packing(Purge valve seal (P/N: 885-2920)
NOTE
: Recap after use the seal remover PV. Keep the seal
remover PV in an accessory box of Chromaster 5160
Pump.
(1) Loosen the purge valve knob (in "Open" direction) and pull it
out.
(2) Screw the seal remover PV into a packing of a valve as the
figure below indicates. Pull down the seal remover PV, and
then pull up the packing to take it out.
(3) Attach a new valve seal at the end of drain knob shaft.
(4) Insert the drain knob into the valve unit and secure it.
Seal remover PV
Cap
Purge valve knob
Purge Valve Seal
Seal remover PV
New purge valve seal
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...