6-2-(105)
Normal Serial Interface
Section 6-2
6-2-3
Output Format (Normal)
When Normal data is set to a unit, the measurement results are output by out-
put number in ascending order. Measurement results are output using normal
communications when the communications mode is set to menu operation.
Outputting ASCII Data
■
Output Format
<Data 0 measurement>, <Data 1 measurement>, - - -
<Data 7 measurement> Delimiter
CHECK
Settings such as the data format, number of digits, and data delimiter can be
changed as necessary. The following table shows the default settings.
SeeAlso
Refer to the
1RUPDO'DWD
.
■
Condition Settings
Example
This example shows an example data output.
CHECK
The field separator is required in order for the output to continue to the next
data field.
• Digital Output Range
-9,999,999.999
measurement
9,999,999.999
The minimum value of “-9,999,999.999” will be output for measurements
less than -9,999,999.999.
The maximum value of “9,999,999.999” will be output for measurements
greater than 9,999,999.999.
• JG (Judgement) Outputs
The following values are output when JG (Judgement) has been set:
OK: 0
NG: -1
CHECK
After measurements have been made in Run Mode, the data up through the
last measurement will be output even if the mode is changed to another mode.
The data output will not be interrupted midway.
Item
Setting
Output format
ASCII
Digits of integer
8 digits
Digits of decimal
3 digits
Minus
-
Field separator
Comma
Record separator
Delimiter
0 Suppress
OFF (Spaces are inserted in empty digits.)
12345.678 , 567.321 , - 76.912 , ------------ , 65987.165
Data 0
measurement
Data 1
measurement
Data 2
measurement
Data 7
measurement
Field separator
Record separator
Delimiter