2-12-(14)
Detecting Binary Defects
Section 2-12
A cursor will appear at the position of the center of gravity.
2.
To change the position, use the Up/Down and Right/Left Keys to move
the cursor.
3.
Press the ENT Key.
The setting will be registered and the screen in (1.) will return.
STEP 5: Setting the Judgement Conditions
Make settings for the area and center of gravity.
1.
Select Judgement conditions.
251,184
Reference registration
ENT:Set :Move
+
Reference position
Measurement region
Judgement conditions
End
Area range (0 to 9,999,999.999)
Range of movement of the measurement object
in the X direction (-9,999.999 to 9,999.999)
Range of movement of the measurement object
in the Y direction (-9,999.999 to 9,999.999)
: Measurement results for the displayed image
Use these values as a reference for setting upper and lower limits.
Range for an OK judgement
Area : 2035.000
[ 0.000 :
247808.000
]
Gravity X : 180.000
[ 0.000 : 511.000]
Gravity Y : 250.000
[ 0.000 : 483.000]
Region0(Gravity and area)
Binary
Measurement conditions
Region
Reference registration
Judgement conditions
Detailed conditions
Clear